35040 Advanced Therapy Dosimeter, TNC-F/M

35040 Advanced Therapy Dosimeter, TNC-F/M
35040 Advanced Therapy Dosimeter, TNC-F/M
Item Number :
321-006

35040 Advanced Therapy Dosimeter, TNC-F/M.

Details

Open section P-Diodes_Dosimeters_Electrometers(PDF) to view related products. (Opens in a new tab)

35040 Advanced Therapy Dosimeter, TNC-F/M.

The 35040 Advanced Therapy Dosimeter is a reference grade instrument used to measure the charge and current from ion chambers in Radiation Therapy, and provides bias voltage for all commonly used chambers.  The ATD large clear display offers direct readings of charge, current, time and radiation units over a wide range.  The user can customize the display for basic use or for specialized applications such as brachytherapy.  The 35040 Advanced Therapy Dosimeter exceeds the recommendations of calibration laboratories for leakage, linearity, and stability by a wide margin.

• Now with timer function and wider dynamic range

• Ultra long-term stability error of approximately 0.1% per five years.

• Virtually removes effects of system leakage during measurement. Uncorrected leakage <10 fA over temperature range

• Eleven user-defined bias settings from -500 V to +500 V

• Read out in C, A, R, Gy, Sv, Bq, and more

The 35040 Advanced Therapy Dosimeter, provides the long-term stability and accuracy demanded for calibrations, quality assurance programs, and protocols in Radiation Therapy.

The 35040 Advanced Therapy Dosimeter is fully stable within five minutes.  In battery operation or using AC Line, the ATD measures dose and effective exposure time in a single exposure, thus eliminating the need for multiple exposures for Cobalt 60 and brachytherapy measurements.  Front panel controls select ion chamber calibration factors, facilitate entry of temperature and pressure values for air density correction, allow bias voltage selection, threshold level, timer control, and choice of display screens.  The user-customized display screens can simultaneously show dose, exposure time, dose rate, effective exposure time, average current/rate, accumulated charge/dose, bias voltage, leakage, and other important information that ensures the validity of the instrument.

The customization software allows design of 16 screens that display conditions, parameters, values and text.  Up to 32 chambers factors, 11 bias voltages can be programmed.  It is PC compatible and connects via a RS-232 cable.

Features
• Wide measurement range, up to 1,000 µA and 19.999 mC for HDR Brachytherapy applications

• Automatic reset and hold of measured valued between exposures

• Front panel adjustment of exposure threshold and user disable of threshold to permit manual operation

• Thirty-two ion chamber calibration factors

• Automatic power-down after user-specified time period, when operating from battery supply

• Annunciators warn of low battery, low bias, and operational errors

• Large capacity battery provides eight hours of continuous operation; fast recharge in less than three hours, even during operation

• AC line operation over the range 100 to 240 VAC and 47 to 63 Hz without operator intervention

• Charge and current calibration factors entered by calibration laboratories at user’s option

• Front and rear panel ion chamber connections

Specifications:
Dose and Rate:
 Display: See picture above
Effective Exposure Time Ranges:  See picture above
Stability:  Designed for ultra long-term stability error of approximately 0.1% per five years.
Leakage:  Virtually removes effects of total system leakage during measurement.  Uncorrected leakage <10 fA over temperature range.
Linearity:  Maximum non-linearity variation from straight line of 0.1% of all charge and current ranges.
Resolution:  0.005% of range (4.5 digits) for charge, dose, average rate and average current; 0.05% of range (3.5 digits) for current and rate.
Warm-up:  Fully meets specifications within five minutes of   applying power.

Measurement Accuracy:  18° C to 28° C (64° to 82° F)
Charge ± (0.20% readings plus two counts)
Current ± (0.20% reading plus two counts)

Bias:  Eleven (11) user-programmable steps from -500 V to +500

V in 0.1 volt increments.  Accuracy ±0.3 volt for loads <0.2 mA.  Front panel selectable.

Ion Chamber Calibration Factors:  Thirty-two (32) user- programmable calibration factors.  Front panel selectable.
Display Unit:  All practical radiation and electrical units.
RS-232 Computer Configuration:  For customizing and data transfer.
Physical Dimensions:  9.4” W x 10” D x 4.0” H
Weight:  10 lbs.
Power:  Internal Lead Acid Battery; Integral Charger Operates 100 to 240 VAC
(47 - 63 Hz)
Connectors:  Triax TNC-F/M front/rear