CT Electron Density Phantom

CT Electron Density Phantom
CT Electron Density Phantom
Item Number :
682-467

CT ELECTRON DENSITY PHANTOM.

Details

Open section B-CT_Simulators_CT_Products_Lasers(PDF) to view related products. (Opens in a new tab)

CT ELECTRON DENSITY PHANTOM.

Create CT-to-density tables with ease for TPS commissioning.


• Cover a wide range of electron density values

• Help ensure accurate calculations of dose distributions

 

The American Association of Physicists in Medicine1 (AAPM) and International Atomic Energy Agency2 (IAEA) report that Treatment Planning Systems (TPS) convert Hounsfield Units (HUs) to electron densities for accurate calculations of dose distributions. This is commonly done using electron density reference materials that enable the verification procedure.

The only Solid Waterfi Insert Holder
The CT Electron Density Phantom consists of a zero HU Solid Water® disk the size of an average pelvis. Sixteen insert chambers in the disk are designed to be used with interchangeable Tissue Mimicking Materials (TMM), used for characterization. Our standard Electron Density Insert Set
includes 13 different materials with a wide range of electron density values. Use the electron density inserts to create CT-todensity tables needed for TPS commissioning. The Solid Water disk also contains eight, 1mm diameter holes spaced at 50 mm intervals around the center of the phantom to test the geometric accuracy of the CT scanner. The same Solid Water disk can be used for calcium and iodine inserts to evaluate the Dual Energy characteristics of CT scanners.


Specifications

Disk Material: Zero HU Solid Water®
Size: 12.9" Dia x 2" H (33 x 5 cm)
Geometric Accuracy Testing: 1mm diameter holes spaced at 50 mm intervals

Weight
Disk and Standard Inserts: 10 lb (4.6 kg)
Phantom with case: 14.5 lb (6.6 kg)
Warranty: One year

1 American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality Assurance for Clinical Radiotherapy Treatment Planning
2 IAEA TECDOC-1583. Commissioning of Radiotherapy Treatment Planning Systems: Testing for

Tissue Mimicking Insert: Lung (LN-300)

Electron Density Relative to Water: 0.29
Physical Density g/cm3: 0.30

Tissue Mimicking Insert: Lung (LN-450)
Electron Density Relative to Water: 0.40
Physical Density g/cm3: 0.45

Tissue Mimicking Insert: Adipose (AP6)
Electron Density Relative to Water: 0.90
Physical Density g/cm3: 0.92

Tissue Mimicking Insert: Breast
Electron Density Relative to Water: 0.96
Physical Density g/cm3: 0.99

Tissue Mimicking Insert: Zero HU Solid Water (x4)
Electron Density Relative to Water: 099
Physical Density g/cm3: 1.02

Tissue Mimicking Insert: Brain

Electron Density Relative to Water: 1.05
Physical Density g/cm3: 1.05

Tissue Mimicking Insert: Liver (LV1)
Electron Density Relative to Water: 1.07
Physical Density g/cm3: 1.08

Tissue Mimicking Insert: Inner Bone
Electron Density Relative to Water: 1.09
Physical Density g/cm3: 1.12

Tissue Mimicking Insert: Bone (B200)
Electron Density Relative to Water: 1.11
Physical Density g/cm3: 1.15

Tissue Mimicking Insert: Bone (CB2-30% Mineral)
Electron Density Relative to Water: 1.28
Physical Density g/cm3: 1.34

Tissue Mimicking Insert: Bone (CB2-50% Mineral)
Electron Density Relative to Water: 1.47
Physical Density g/cm3: 1.56

Tissue Mimicking Insert: Cortical Bone (SB3)
Electron Density Relative to Water: 1.69
Physical Density g/cm3: 1.82

Tissue Mimicking Insert: True Water
Electron Density Relative to Water:  1.00
Physical Density g/cm3:  1.00