View PDF Catalog   

Closeout Items   

Fluke CT Ion Chamber, High Sensitivity 10 cc

Fluke CT Ion Chamber, High Sensitivity 10 cc
Fluke CT Ion Chamber, High Sensitivity 10 cc
Item Number :
300-510

Fluke CT Ion Chamber, High Sensitivity 10 cc for multislice CT, with triax BNC: used with the 35040 ATD and other electrometer/dosimeters, including TRIAD™ and TRIAD™ TnT.

Details

Open section B-CT_Simulators_CT_Products_Lasers(PDF) to view related products. (Opens in a new tab)

Fluke CT Ion Chamber, High Sensitivity 10 cc for multislice CT, with triax BNC: used with the 35040 ATD and other electrometer/dosimeters, including TRIAD™ and TRIAD™ TnT.

Specifications
Detector Type:
  Vented air ion chamber
Volume:
  10.1 cc
Sensitive Length:
  10.0 cm
Chamber Material:
  Acrylic (PMMA)
Chamber Outside Diameter:
  0.5 in ± 0.015 in (12.7 mm ± 0.4 mm)
Chamber Inside Diameter:
  0.45 in (11.44 mm)
Chamber Wall Thickness:
  77 mg/cm2
Electrode Material:
  Aluminum, 1100
Sensitivity:
  3.2 R•cm/nC (nominal) or 0.3/nC
Standard Calibration:
  100 kVCP, 5.5 mm Al HVL (NIST Tech. M100)
Response Uniformity Along Axis:
 ± 3% over central 90% of active length
Beam Orientation:
  Normal to chamber axis
Leakage Current:
  (300 V collection potential)  Less than 10-14A at 10 min polarization time
Intensity Limits: Continuous beam:
  31.6 R/Sec, (1% recombination loss)
Pulsed Beam:
  15.8 mR/pulse (1% recombination loss)
Collection Time:
  0.478 mSec
Cable Length:
  3' (0.9 m)
Operating Voltage:
  -300 V