Specifications
Detector Type: Vented air ion chamber Volume: 3.2 cc Sensitive Length: 10.0 cm Chamber Material: Polystyrene Chamber Inside Diameter: 6.4 mm Chamber Wall Thickness: 54 mg/cm2 Electrode Material: Aluminum Sensitivity: 10 R•cm/nC (nominal) Standard Calibration: 100 kVCP, 5.5 mm Al HVL (NIST Tech. M100) Response Uniformity Along Axis: ± 3% over central 90% of active length Beam Orientation: Normal to chamber axis Phantom Adapter OD: 1.27 ± 0.04 cm (0.50 ± 0.015 in) Leakage Current (300 V collection potential): Less than 1013 A at 10 min polarization time, less than 1014 A at 2 hr polarization time Intensity Limits: Continuous beam: 4.86 kR/min, Pulsed Beam: 51.5 mR/pulse (1% recombination loss) Maximum Pulse Repetition Rate: 3.3 kHz Cable Length: 3 ft (0.9 m) Operating Voltage: -300V
|