Atomlab 100 Dose Calibrator

Designed for facilities receiving unit doses for PET

Item #

Description

994-080

Atomlab 100 Dose Calibrator with RS-232 Port
Includes Vial/Syringe Dipper and Well Insert

994-016

Dose Calibrator Interlocking Shield Rings, 2” Lead

994-082

Vial Moly Shield, 0.28” Lead

994-083

Vial Moly Shield Syringe, 0.28 Lead

994-085

Source, Reference Set, Includes: Co-57, 5 mCi; Cs-137,
200 µCi; Ba-133, 250 µCi. Calibrated to ± 5%

994-086

Lineator

994-087

Replacement - Vial/Syringe Dipper

994-088

Replacement - Well Insert

Features

  • All functions microprocessor controlled
  • Ultra-fast response
  • Automatic range selection Ranges up to 9.999 curies of Tc-99m or 2.5 curries of F-18
  • Pre-programmed for the most commonly used radionuclides
  • Display in Curies or Becquerels
  • Automatic background subtraction and zeroing at the touch of a button
  • Remote ionization chamber with .25” (.64 cm) lead shielding and 10-foot cable
  • Self-diagnostic software
  • Daily constancy isotope keys
  • Electronic power supply (no battery in chamber)

The Atomlab 100 provides fast, accurate radionuclide activity measurements with performance that easily surpasses the most stringent regulatory requirements.

The unit is simple to operate. There are 13 isotopes selection keys, ten are pre-programmed for the most commonly used radionuclides, and three are user defined. There are 88 isotope - specific dial values listed in the manual, including Y-90 and Sr-89. Any key can be reprogrammed by the user for a desired isotope.

Activity is displayed on a LED readout in either Curie or Becquerel units. Background correction and zero adjustment are performed at the touch of a button. Range selection is automatic.

Activity measurements are performed by a microprocessor controlled electrometer located within the detector assembly of the ionization chamber. The chamber is shielded with .25” (.64 cm) lead. It can be located up to ten feet away from the display unit. Chamber bias is generated within the display unit by an electronic high voltage supply, eliminating the need for expensive battery changes.

The RS-232 port enables the Atomlab 100 Dose Calibrator to communicate with most commercially available nuclear medicine management systems.

Specification:

Isotope Selection Keys: Ten pre-programmed: Tc-99m, T1-201, Mo-99, I-123, Xe-133, Ga-67, In-111, I-131, Cs-137, and Co-57; Three additional keys for user-defined isotopes.
Activity Range: 0.01 µCi to 9999 mCi (.001 mBq to 399.9 GBq) of Tc-99m
Energy Range: 25 KeV to 3 MeV photons

Response Time: One second for doses greater that 2 mCi; Three
seconds between 200 µCi and 2 mCi; 3-30 seconds below 200  µCI
Detector Linearity: ±1% or 0.2 µCi, whichever is greater
Electrometer Linearity: ± 1% or 0.2 µCi, whichever is greater
Electrometer Accuracy: ± 1% or 0.2 µCi, whichever is greater
Overall Accuracy: ± 3% or 0.3 µCi, whichever is greater. Overall
accuracy is affected by such factors as the accuracy of the  specific source calibration, geometric variations due to sample  volume or configuration, detector linearity, electrometer accuracy  and readout accuracy.
Repeatability: ± 0.3% above 1 mCi short term (24 hr); 1% long  term (1 yr); exclusive of background.
Digital Calibration Dial: Four-digit LED dial display with  increment/decrement keys to change the value. Range is from  0.1 to 999.
Detector: Well-type pressurized ionization chamber, with Argon  fill gas. 
Well Opening: 2.75” (7 cm), well depth is 10.5” (26.7 cm)
Detector Shielding: .25” (6.3 mm) lead on all sides except top  well opening. Supplementary shielding available.
Chamber Bias: 340 volt electronic power supply
Environmental Operating Conditions: Temperature: 0-40° C. Humidity: 0-90% rh, non-condensing
Power Requirements: 100 to 120 VAC @ 1/2A; 200 to 240 VAC
@ 1/4A
Line Frequency: 50/60 Hz; Detachable line cord; Built-in EMI filter and transient suppression.
Detector and Interface Cables: 10’ (304.8 cm) long, six  conductor cables (two carry power, one is chassis ground, three  carry serial data for digital I/O); Ends terminated with AMP Mate- n-Lock connectors, white with hooded strain relief.

Display Unit:
Dimensions: 12” W x 14.3” D x 3.75” H (30.5 x 36.3 x 9.5 cm)
Weight: 6 lb (2.7 kg)

Detector Unit: 
Dimensions: 7.5” W x 7.5” D x 16.25” H (19 x 19 x 41.3 cm)
Base Dimensions: 4” thick x 12” x 12” (10.16 x 30.48 x 30.48 cm)
Well I.D.: 2.75” dia. x 10.5” H (7 x 26.7 cm)
Well I.D. with Liner: 2.5” dia. x 10.25” H (6.35 x 26 cm)
Lead Shielding: .25” thick (.63 cm)
Weight: 35 lb (16 kg)
Approvals: ETL to UL 3101-1 and cETL to CAN/CSA C22.2 No. 1010.1-M92.