Brachytherapy QA Phantom

Perform QA on sidefire transrectal probes

Item #

Description

906-450

Brachytherapy QA Phantom

  • Internal gid assessment
  • Probe retraction step assessment
  • Volume verification

The Brachytherapy QA Phantom was designed for transrectal ultrasound QA and calibration of brachytherapy systems. It contains targets to assess volume measurements, internal grid accuracy, and probe retraction accuracy.  When scanning towards the top of the phantom, a partial grid of wires appears.  These wires should line up with the grid that appears on your screen thus ensuring correct vertical and horizontal distance measurements.

Five cross wires are embedded within the phantom to determine if the probe is being retracted the specified distance.  Turn the probe 60° to the right or left to visualize and measure the volume of three different calibrated objects, one of which is non-spherical.
Specifications:

Material:  Zerdine™(1)
Attenuation Coefficient: 0.50 ± 0.05 dB/cm/MHz
Speed of Sound:  1540 m/s ± 6 m/s

Targets:
Internal Gid Assessment:
Material:  Nylon Monofilament
Diameter:  0.5 mm
Number of Targets: 13
Position:  B1 - B5, C4, D3, F1 - F5

Probe Retraction Assessment:
Material:  Nylon Monofilament
Diameter:  0.5 mm
Number of Targets: 5
Spacing: 1. cm, 1. cm, 1.5 cm and .5 cm
Position:  Along row 4

Volumes:
Material:  Zerdine™(1)
Sizes: Approximately 4 cc, (S), 9 cc (M) and 23 cc (L). Exact volumes provided on certification sheet
(1) US Pat. # 5196343

Radiation Products Design, Inc     763-497-2071    Fax 763-497-2295     www.rpdinc.com          sales@rpdinc.com