CRC®-15BT BRACHYTHERAPY DOSE CALIBRATOR FOR HDR AND LDR

Item #

Description

344-030

CRC®-15BT Dose Calibrator

344-031

RS-232-R Serial Port

344-033

CRC®-AD15 Auxiliary Display for CRC-15BT

344-035

Cs-137 Seed Holder and Tube Set for CRC-15BT

344-036

Cs-137 Tube Only

344-037

Ir-192 Ribbon Source Holder for CRC-15BT

344-038

I-125 Linear Seed Tube and Holder Set for CRC-15BT

344-039

I-125 Seed Tube Only

344-040

I-125 Seed Cartridge Holder for CRC-15BT

344-041

Mick Cartridge Holder CRC-15BT

994-027

Okidata Printer

994-028

Epson Roll Printer

994-029

Epson Ticket Printer

Source Holders for 15BT Dose Calibrator

Features:

  • Accurately and rapidly reads both HDR, LDR and IVBT sources
  • Readout in Ci, or Air Kerma strength
  • Pressurized chamber for enhanced sensitivity which means no temperature or pressure corrections necessary
  • Built-in auto-calibration, QC programs, background subtraction and system test
  • Very simple and easy to use, offering a wide variety of features
  • Multiple seed/seed train measurement capability
  • Specialized holders for seed trains and other source geometries

Both HDR and LDR brachytherapy sources are quickly and precisely measured by the CRC®-15BT.

The console of the CRC®-15BT is designed for ease and efficiency. There are 14 nuclide keys that can be programmed to meet the users needs.  The unit has a large LCD display that will show the isotope name, activity (in Ci or U) and calibration number.  A custom alphanumeric keyboard drives all of the menus and functions.

The chamber of the CRC®-15BT has 1/8” (3.2 mm) mechanically reinforced lead shielding.  The chamber is a sealed thin wall, deep well design with 2.5 atmospheres of Argon gas and a 500 V bias in order to achieve optimal response time (within 2 seconds) without temperature or pressure corrections.

ADCL Calibration is available with the CRC®-15BT to ensure maximum accuracy.  Calibrated with the specific brachytherapy source(s) utilized in your department, activity can be measured precisely in either Curies (Ci) or Air Kerma strength (U).

The CRC®-15BT software is a menu driven design that will allow the user to move rapidly through all of the system functions.  The Daily QC protocol will perform zero adjust, background subtraction, system test, accuracy (with a deviation percentage) and constancy.  The nuclide library is programmed with more than 80 nuclides and half-lives, plus available space for 10 additional nuclides.  The system memory can also support up to 20 brachytherapy sources with isotope name, calibration number, Air Kerma strength and ADCL factor.

With optional features such as the RS-232 computer interface port, a selection of printers (ticket, roll or full page) or the CRC®-AD15 auxiliary display you can configure a system that is right for your department.

Specifications

Ionization Chamber
2.5 atmosphere ultra pure Argon gas
500 Volt bias

Measurement Range: 1mCi to 12 Ci

Display Screen
Type:  Dot Matrix liquid crystal display
Format:  Direct reading in U or Ci
U/Ci reading:  User selectable
Values Displayed:  Seed/Source name, ADCL factor, calibration number

Electrometer
Accuracy:  Better than ±2%
Linearity: Within ±2%
Response Time: Within 2 seconds, 4 to 16 sec for very low activity samples, (user selectable)

Overall Accuracy - Accuracy Determined by:
1. Calibration for the specific Brachytherapy source and the sample configuration
2. Accuracies of standard sources used for calibration of electrometer

Repeatability of Measurement
Repeatability: ± 1% within 24 hours during which time the calibrator is continuously powered on

Power Requirements:  90-127 VAC (50/60 Hz) 0.1A; 180-250 VAC (50/60 Hz) 0.05A
Display Unit Dimensions: 5” x 10” x 10”
Weight:  4 lbs
Chamber Unit Dimensions: 12” H x 7” dia
Well Diameter: 2.4”
Weight:  27 lbs