DOSE-1 ELECTROMETER

Item #

Description

321-100

DOSE-1 Electrometer

The Dose-1 is a high-precision reference class electrometer that significantly exceeds the recommendations of the IEC 60731 and the AAPM ADCL’s.  It is suitable for the use with ion chambers, semiconductors and diamond probes. The DOSE-1 is ergonomically designed and intuitively easy to use.  The standard DOSE-1 connects to either TNC or BNC connector types. Optionally , the DOSE-1 is compatible with TNC, BNC, M-type and BNC/banana connector types.

•  Large, high contrast graphic electro luminescent display with a wide viewing angle for complete and comprehensive display of all measured values, chamber and correction factors on the main screen.
•  Dose, dose rate, average rate, charge, current and dose per monitor unit are all measured and displayed simultaneously in the expanded screen.
•  Time base for the dose rate is selectable prior to or after the measurement.  The selectable units are: second, minute and hour.
•  The operator interface is ergonomically designed and intuitively easy to use with pop-up menus, operated by soft keys.
•  Storage of up to 40 detector specific data sets in a sensor library, including physical and geometrical parameters.
•  Built-in electrical check source as well as leakage and bias voltage testing to ensure maximum reliability and accuracy.
•  The DOSE-1 has a wide range power supply for automatic adoption to mains voltage worldwide. Battery operation with 4 D cells or NiCad rechargeable batteries is optionally available.

Specifications

Bias Voltage:  ±600 V, programmable in steps of 1 V
Sensor Connector Types

Standard Option: Triaxial TNC (threaded) or triaxial BNC (bayonet)
Alternative (“Convertible Option”):  PTW-M, BNC/Banana and triaxial TNC, triaxial BNC
Temperature Range: 15 - 35°
Relative Humidity: 10 - 80% rel humidity, max. 20 g/cm3
Power Supply (Mains):  100 - 240 V, 50/60 Hz
Battery Operation (Optional):  4 D-type NCd / NMH - accumulator cells
PC Interface:  Bidirectional RS-232, configuration software
Outer Dimensions:  6.5” H. x 10.2” W. x 10.2” D.
Weight:  1.2 lbs.

Measuring Modes/Range

Charge (Dose): 40pC to 1.0C at 0.1pC resolution
Current (Dose Rate): 40 pA to 1000nA at 0.1pA resolution

Measuring Quantities and Units

Electrical:  charge (C), Current (A)
Integrate: Gy, Sv, R, rad, rem
Time Base for Rate:  Second, minute, hour
Interval Time Range: 1 to 9999 sec
Accuracy/Repeatability:  ±0.2%
Leakage Current: >±10fA, typically 1fA
Linearity: <±0.25% in whole range
Stability: <±0.25% per year
Display: Graphic electro luminescent, 160° viewing angle
Zero:  Automatic, within 60 sec.
Memory:  All setup and detector parameters stored in EEPROM
Background Compensation: On/off mode with memory

Radiation Products Design, Inc     763-497-2071    Fax 763-497-2295     www.rpdinc.com          sales@rpdinc.com