DOSE DRAWING SYSTEM

For drawing FDG F-18 from a vial into a syringe

Item #

Description

994-465

Dose Drawing System, PET

994-466

Syringe Shield, Dose Drawing, PET, 5 cc

994-467

Vial Shied, Dose Drawing, PET, 30 ml

994-468

Dose Preparation Stand

If transported in a PET Vial Shipping System, the shipped Vial Shield is placed in the stand and secured with the tightening knob.

Vial Shield shipping cover is replaced with the drawing shield cover.

Syringe is placed in the Dose Drawing Syringe Shield. The Vial Shield roates to 45-degree downward angle. Swing open the cover for Vial Septum access.

Syringe Shield slides upward, allowing the syringe to puncture the Vial Septum. Dose is drawn.

The Dose Drawing Sytem is used to draw FDG F-18 doses from a vial. It consistes of a specially designed Dose Drawing Syringe Shield, Vial Shield and stand.

When drawing FDG F-18, the Vial Shield and Syringe Shield remain on the stand. The stand allows the Vial Shield to rotate and has a fixed stop at the 45-degree downward angle. The Vial Shield is constructed of 1” lead encased in steel. The shield’s top is threaded to allow quick insertion and removal of vials. The top cover incorporates a pivoting shield for septum access that allows the use of a vent needle, if desired, along with the needle from a 5 cc syringe.

The Syringe Shield mounts on a rotating platform that slides the syringe into the vial and locks in position. The dose can then be drawn with forceps to minimize hand exposure. The Syringe Shield is constructed of .75” lead encased in steel. A high-density lead glass window allows viewing up to the 5 cc mark on a 5 cc syringe. 

The PET Dose Drawing System provides a safe and shielded environment while making it easy to draw a FDG F-18 dose.

Specifications

Dose Drawing Syringe Shield
Dimensions: 2.25” dia. x 5.5” L. (5.7 x 14 cm)
Lead Shielding: .75” Thick (1.9 cm)
Lead Glass Window: .25” Thick (.64 cm), 5.6 density

Construction:
Exterior: Steel
End Cap Mechanisim: Stainless Steel
Finish: Powder coat
Weight: 13 lb (5.9 kg)

Dose Drawing Vial Shield
Dimensions: 6.1” H 4” Dia. (15.5 x 10.2 cm)

Lead Shielding:
Cylinder: 1” Thick (2.54 cm)
Top End Cap: 1” Thick (2.5 cm)
Cover: .75” Thick (1.9 cm)

Construction:
Exterior: Steel
Interior: Molded lead
Top End Cap: Aluminum
Bottom End Cap: Plastic
Pivoting Shield: Lexan wear surface on a steel and lead cover.
Finish: Powder coat
Weight: 20 lb (9 kg)

Dose Preparation Stand
Dimensions: 12.6” W x 10.6” D. x 12.6” H (32 x 26.9 x 32 cm)
Construction: Steel
Finish: Powder coat
Weight: 21 lb (9.5 kg)