GAFCHROMIC® EBT FILM

Item #

Description

115-015

Gafchromic® EBT, 8” x 10” (25 sheets/box)

115-017

Gafchromic® EBT-1417, 14” x 17” (10 sheets/box)

GAFCHROMIC® DOSIMETRY MEDIA SELF-DEVELOPING DOSIMETRY FILMS

Gafchromic® EBT self-developing dosimetry film is targeted to applications in external beam radiotherapy, in particular intensity modulated radiation therapy (IMRT) and brachytherapy.
EBT Features:
•  Sensitive to doses ranging from about 1cGy to 800cGy
•  Self-developing in real time
•  Uniform and consistent
•  Faster and lower post-exposure density growth
•  Energy-independent dose response from KeV to MeV
•  Near tissue-equivalent
•  High spatial resolution
•  Insensitive to light
•  Uniformity better than 1.5%
•  Water resistant- usable in water phantoms
•  Digitized with Vidar, Epson, and Microtek scanners
•  EBT sleeves for Vidar scanners are available upon request.

Energy Independence of Gafchromic® EBT Dosimetry Film
Gafchromic® EBT radiochromic film has been designed to measure absorbed dose from high-energy radiation used in IMRT. The response to photons has been found to be energy-independent in the MeV range and measurements at lower energies show that the response changes by only a few percent, even down to 30keV.

The response of Gafchromic® dosimetry film is dependent upon factors specific to the facility such as the energy spectrum of the radiation source, the characteristics of the densitometer, film scanner or other means of measurement, and the elapsed time between exposure and measurement etc. Therefore, the user should determine the dose-density response under the conditions pertaining to the user’s facility and measurement protocol. Since the absolute response of Gafchromic® EBT dosimetry film may vary from lot to lot, the user should re-measure response when changing to a new lot number.

Dose Fractionation
The results in the table below indicate that the response of EBT film exposed to a single fraction is indistinguishable from the response of the film given fractionated exposure and establishes EBT film as an effective dose integrator.

Exposure

Net Visual Density Change
(avg. of 9 measurements)

5Gy in a single fraction

0.681 ± 0.015

5Gy in 1Gy fractions at 30min. intervals

0.677 ± 0.009

Response of Gafchromic® EBT Dosimetry Film to Fractionated Exposure

Radiation Products Design, Inc                 www.rpdinc.com                     sales@rpdinc.com
 763-497-2071    Fax 763-497-2295     5218 Barthel Industrial Drive  Albertville, MN 55301