Electronic Pocket Dosimeter (EPDS)

Item #

Description

343-112

PDM-112 Electronic Pocket Dosimeter

343-117

PDM-117 Electronic Pocket Dosimeter

K343-112-A

K343-117-A• No Reader required
• Measuring range o 1 to 9999 µSv
• Low battery/ overflow indication
• High sensitivity
• Easy-to-read digital display
• Simple operation
• Sensitive to gammas and x-rays 20 keV and above

The Electronic Pocket Dosimeter uses semiconductor detectors that are sensitive to gammas and x-rays. Each dosimeter indicates the integrated radiation dose equivalent on a 4-digit liquid crystal display.
[Note: 1 µSv= 0.1 mRem]

Specifications for 343-112

Radiation Detected: Gamma (X-ray)
Energy Threshold: 40 keV
Detector: Silicon semiconductor
Measurement Range: 1 to 9999 µSv
Energy Dependence: Within ±30% from 50 keV to 3 MeV
Accuracy: Within ±10% from 10 to 9999 µSv (calibrated by 137Cs using a slab phantom.)
Linearity: Within ±10% up to 0.1 Sv/h ; within ±20% from 0.1 to 0.3 Sv/h
Display: 4-digit LCD
Other: Display test function, overflow indication, low battery indication, data storage function*
Power Supply: One lithium battery, CR2450 (coin size)
Battery Life: Approx. one month of continuous use
Environmental Requirements: Temperature 0 to +45°C, Relative Humidity up to 90% (non-condensing)
Dimensions: Approximately 1.2”W x 5.7”H x 0.5”D excluding the clip.
Weight: Approximately 1.75 oz.
*The measured value is stored unless the battery is removed or the power switch is pressed for ten seconds when turning on the instrument.

Specifications for 343-117

Radiation Detected: Gamma (X-ray)
Energy Threshold: 20 keV
Detector: Silicon semiconductor
Measurement Range: 1 to 9999 µSv
Energy Dependence: Within ±30% from 30 keV to 200 keV
Accuracy: Within ±20% from 10 to 9999 µSv (calibrated by 40 keV of X-ray using a slab phantom.)
Linearity: Within ±10% up to 30 mSv/h ; within ±20% from 30 to 100 mSv/h
Display: 4-digit LCD
Other: Display test function, overflow indication, low battery indication, data storage function*
Power Supply: One lithium battery, CR2450 (coin size)
Battery Life: Approx. half month of continuous use
Environmental Requirements: Temperature 0 to +45°C, Relative Humidity up to 90% (non-condensing)
Dimensions: Approximately 1.2”W x 5.7”H x 0.5”D excluding the clip.
Weight: Approximately 1.75 oz.

*The measured value is stored unless the battery is removed or the power switch is pressed for ten seconds when turning on the instrument.

Note: As an attempt to verify operation or accuracy, never irradiate the dosimeter directly to a source such as X-ray as it is not designed for exposures in locations where doserate is 100 mSv/h or more. For questions concerning calibration or repair, please call 1-800-497-2071. It is recommended by the manufacturer that dosimeters should be tested for accuracy on an annual basis by a qualified lab. Units failing to meet specifications cannot be repaired or re-calibrated.