• Hemispherical geometry for easy placement and alignment • High impedance for minimum drift • High output - 32nC/Gy typical • Excellent output uniformity vs. gantry angle • Outstanding stability - <0.1%/kGy at 6MV photon beam • Available in various water equivalences for photons and electrons • Vinyl-sheathed cable eliminates kinking
The EquiDose II diode detectors are a new generation of solid state detectors exhibiting superior characteristics. Radiation damage is extremely low, with detectors typically retaining better than 94% of their original response after 15,000 exposures of 2Gy (3kGy). Axial directional response variation is typically less than 1%, ±90° from perpendicular. A low temperature coefficient minimizes sensitivity variation due to patient heat transfer during treatment.
Each photon EquiDose II detector responds most effectively over one of three energy ranges. When used within its stated energy range, a calibrated detector will register a dose approximately equal to that expected at dmax. They operate in shorted junction mode and require no bias voltage. Specifications
Detector type: n type diode Detector size: 0.8 x 0.8 mm2 Effective Detection Thickness: 50µm Active Volume: 0.136mm3 Nominal Output: 32 nC/Gy Input Impedance: >200MW at 10V @ 24°C Temp. Coefficient: <0.5% / °C Linearity: Within 0.2%, 50 to 500 rads Output Stability: 2% loss of sensitivity at 1kGy(10MeV electron beam), 0.1 % loss of sensitivity at 1kGy (6MV photon beam) Directional Dependence: optimized for each model Construction: Hemispherical with 12.7 mm flat bottom Cable: 2m low-noise coax, 2.5mm thick Connector: BNC (Lemo optional) Note: Older Models Lemo-M, Lemo Adapter Recommended
|