• Hemispherical geometry for easy placement and alignment • High impedance for minimum drift • High output - 32nC/Gy typical • Excellent output uniformity vs. gantry angle • Outstanding stability - <0.1%/kGy at 6MV photon beam • Available in various water equivalences for photons and electrons • Vinyl-sheathed cable eliminates kinking
The EquiDose II diode detectors are a new generation of solid state detectors exhibiting superior characteristics. Radiation damage is extremely low, with detectors typically retaining better than 94% of their original response after 15,000 exposures of 2Gy. Axial directional response variation is typically less than 1%, ±90° from perpendicular. A low temperature coefficient minimizes sensitivity variation due to patient heat transfer during treatment.
Each photon EquiDose II detector responds most effectively over one of three energy ranges. When used within its stated energy range, a calibrated detector will register a dose approximately equal to that expected at dmax. They operate in shorted junction mode and require no bias voltage.
Specifications
Detector type: n type diode Detector size: 0.8 x 0.8 mm2 Sensitivity: 32 nC/Gy Input Impedance: >200MW @ 24°C, 10MV reverse bias Temp. Coefficient: 0.5% / °C Output Stability: 2% loss of sensitivity with10MeV electron beam, 0.1% loss of sensitivity at 1000G, 6MV photon beam Construction: Hemispherical with 12.7 mm flat bottom Cable: 2m low-noise coax, 2.5mm thick with Tygon® Sheath Connector: BNC-M (Lemo optional)
|