Repair Services for the 3M #6586 Fletcher-Suit Delclos Shields Afterloading Applicators |
||||||||||||||
![]() |
||||||||||||||
|
Radiation Products Design does fiber optic inspection of tungsten shielding inside of the ovoid. We can detect if the tungsten shields are mounted at the top and the bottom of the ovoid. |
||||||||||||||
|
FDA is using this notice to alert users of Fletcher Suit applicators of this potential problem. FDA believes that all Fletcher Suit applicators should be inspected to determine whether or not the closed end of the colpostats or the colpostats themselves are adequately secured, and that any applicators that are questionable should not be used. Until recently, manufacturers guidance on cleaning and sterilization may not have been clear. The applicators should be cleaned using only a mild detergent and water, and then thoroughly rinsed with water before sterilization. Although, to date, FDA has only received reports of the problems with Fletcher Suit applicators, it is also recommending that all applicators of similar design be checked, and that the same cleaning instructions be followed.” The above article appears in the American Association of Physicists in Medicine. Volume 14 No. 3, May/June 1988, Pages 7 and 8. |
||||||||||||||
|