LINEAR 5IVE MOSFET ARRAY

Item #

Description

322-360

Linear 5ive MOSFET Array, Standard Sensitivity

322-362

Linear 5ive MOSFET Array, High Sensitivity

322-364

Linear 5ive MOSFET Array, Extreme Sensitivity

The Linear 5ive MOSFET Array is ideal for LDR / HDR Brachytherapy and IMRT QA.
The Array contains five isotropic, energy independent MOSFET dosimeters, at 2cm intervals. The MOSFETs on the array can be read simultaneously, facilitating multiple dose measurements at several spatial points. The radio-opaque marker, located at the tip of the array, enables visualization under X-Ray imaging (CT, Fluoroscopy), allowing for five dosimetry points to be accurately located and easily compared to target doses.

Dynamic Dose Measurements for Brachytherapy Dosimetry
The Linear 5ive MOSFET Array™, when used in combination with the mobileMOSFET™, is the first and only commercially available combination that allows for real-time quality assurance of all brachytherapy procedures, without a significant investment in extra time.

Real-time dose profiling is provided by the in-vivo use of a mobileMOSFET Wireless Dosimetry System. the Linear Array is used for various HDR applications such as prostate and gynecological brachytherapy using Item 322-360.

When placed on the surface of the breast, the array can also validate Mammosite™ treatments. Implant and LDR dosimetry is performed using the higher sensitivity model, Item 322-362.

When inserted directly into a urethral catheter, the dose results provide immediate assessment of post-implant base and apex dose coverage, as well as the dose to organs at risk such as the urethra, rectum or bladder. This real-time dose feedback allows assessing the quality of the seed implant program in LDR and HDR brachytherapy. Absolute dose measurements or dose rate measurements are obtained in real-time. This helps to validate the quality of the treatment, and ultimately the quality of life for the patient.

Features
• Suitable for in-vivo dosimetry and beam QA
• For use in both electron and photon modalities
• Isotropic response for 360° gantry angles
• Dose rate and energy independent
• Visible under CT or Fluoroscopy imaging
• Waterproof, resistant to body fluids and liquid sterilization
• Small, lightweight, flexible, and rugged for extensive handling

Compatibility
• mobileMOSFET Dose Verification System
• AutoSense Dose Verification System & Dual Bias Supply for Linear 5ive MOSFET Array
• Not compatible with MOSFET 20 Dose Verification System

Radiation Characteristics
• 20,000 mV lifetime (~20,000 cGy on standard sensitivity setting)
• Five active detection points (0.04 mm2 each)
• Suitable for photon and electron modalities
• Isotropic response (±3% for 360°)
• Temperature independent
• Visible under CT or Fluoroscopy with a radio-opaque tungsten marker at the tip.

Additional Applications
• IMRT IGRT, IORT, QA and In vivo
• Rectal Dose Measurements
• Skin Dosimetry
• Beam Profiling
• Fluoroscopy / CT Dose Verification
• External beam radiotherapy / TBI

Dimensions
1.5 mm W x 46 cm L x 1.3 mm T
2 cm Inter-MOSFET Spacing

Three linear arrays with different sensitivities are offered to accommodate all clinical and research applications:

Linear 5ive Array

Common Use

Standard Sensitivity Bias

High Sensitivity Bias

322-360

HDR brachytherapy, MammoSite

0.98 mV/cGy
(for 192Ir)

1.38 mV/cGy
(for 192Ir)

322-362

LDR brachytherapy

11.1 mV/cGy
(for 125I)

15.2 mV/cGy
(for 125I)

322-364

LDR brachytherapy, diagnostic x-rays

25.8 mV/cGy
(for 125I)

37.2 mV/cGy
(for 125I)

Note: sensitivities noted above are under full build-up (60Co)

All of these arrays continue to yield dose reproducibility at standard sensitivity bias at 1ohm

Linear 5ive Array

20 cGy

200 cGy

322-360

< 2%

< 1.0%

322-362

< 2%

< 1.5%

322-364

< 2%

< 1.5%