LINEAR 5IVE MOSFET ARRAY

Item #

Description

322-360

Linear 5ive MOSFET Array, Standard Sensitivity

322-362

Linear 5ive MOSFET Array, High Sensitivity

322-364

Linear 5ive MOSFET Array, Extreme Sensitivity

The Linear 5ive MOSFET Array is ideal for LDR / HDR Brachytherapy and IMRT QA.

Dynamic dose feedback is critical in determining the quality of all brachytherapy procedures. In HDR, it allows for corrective planning of dose fractions, whereas in LDR it enables the assessment of the quality of the seed implant program. Further real-time, in vivo dose monitoring allows the physician to monitor the dose to critical organs such as the urethra and rectum during brachytherapy procedures. In fact, this has become so important that the American Brachytherapy Society (ABS) recently issued brachytherapy dosimetry recommendations with the following statement:
Ideally, one should strive for on-line, real-time intraoperative dosimetry to allow for adjustment in seed placement to achieve the intended dose [1,2].

The Array contains five isotropic, energy independent MOSFET dosimeters, at 2cm intervals. The MOSFETS on the array can be read simultaneously, facilitating multiple dose measurements at several spatial points. The radio-opaque marker, located at the tip of the array, enables visualization under X-Ray imaging (CT, Fluoroscopy), allowing for five dosimetry points to be accurately located and easily compared to target doses.

Features
• Suitable for in-vivo dosimetry and beam QA
• For use in both electron and photon modalities
• Isotropic response for 360° gantry angles
• Dose rate and energy independent
• Visible under CT or Fluoroscopy imaging
• Waterproof, resistant to body fluids and liquid sterilization
• Small, lightweight, flexible, and rugged for extensive handling

Compatibility
• mobileMOSFET Dose Verification System
• AutoSense Dose Verification System & Dual Bias Supply for Linear 5ive MOSFET Array
• Not compatible with MOSFET 20 Dose Verification System

Radiation Characteristics
• 20,000 mV lifetime (~20,000 cGy on standard sensitivity setting)
• Five active detection points (0.04 mm2 each)
• Suitable for photon and electron modalities
• Isotropic response (±3% for 360°)
• Temperature independent
• Visible under CT or Fluoroscopy with a radio-opaque tungsten marker at the tip.

Additional Applications
• IMRT IGRT, IORT, QA and In vivo
• Rectal Dose Measurements
• Skin Dosimetry
• Beam Profiling
• Fluoroscopy / CT Dose Verification
• External beam radiotherapy / TBI

Dimensions
1.5 mm W x 46 cm L x 1.3 mm T
2 cm Inter-MOSFET Spacing

[1]-S. Nag, W. Bice, K. DeWyngaert, et al. “The American Brachytherapy Society recommendations for permanent prostate brachytherapy post-implant dosimetric analysis”.Int J. Radiat. Oncol. Biol. Phys.,46, 221-230, (2000)

[2]-S. Nag, J.P. Ciezki, R. Cormack, et al. “Intraoperative planning and evaluation pf permanent prostate brachytherapy: report of the American Brachytherapy Society”.Int J. Radiat. Oncol. Biol. Phys., 51,1422-1430 (2001)

Three linear arrays with different sensitivities are offered to accommodate all clinical and research applications:

Linear 5ive Array

Common Use

Standard Sensitivity Bias

High Sensitivity Bias

322-360

HDR brachytherapy, MammoSite

0.98 mV/cGy
(for 192Ir)

1.38 mV/cGy
(for 192Ir)

322-362

LDR brachytherapy

11.1 mV/cGy
(for 125I)

15.2 mV/cGy
(for 125I)

322-364

LDR brachytherapy, diagnostic x-rays

25.8 mV/cGy
(for 125I)

37.2 mV/cGy
(for 125I)

Note: sensitivities noted above are under full build-up (60Co)

All of these arrays continue to yield dose reproducibility at standard sensitivity bias at 1ohm

Linear 5ive Array

20 cGy

200 cGy

322-360

< 2%

< 1.0%

322-362

< 2%

< 1.5%

322-364

< 2%

< 1.5%

Radiation Products Design, Inc         www.rpdinc.com           sales@rpdinc.com
 763-497-2071   Fax 763-497-2295   5218 Barthel Industrial Drive Albertville, MN 55301