The Keithley Model 602 is a high quality, general purpose electrometer capable of measuring low currents, low voltage levels and high resistances. It is best known in the industry for its stability and extremely low input offset current on the order of less than 5 fA.
This electrometer is modified to make is expressly suited for medical dosimetry applications. The modifications include:
- Installation of a 150/300 volt electronic bias supply with reversible polarity
- Replacement of the analog meter with a 4 1/2 digit LCD
- nstallation of a single 12V rechargeable battery with wall-mount
- UL listed transformer/charger
- Installation of circuitry to allow bias voltage and battery condition to be read on the display
- Installation of a metal connector cap with retaining chain
- Color-coding of switch positions for ease of use
- Coulomb scale calibration to ±0.1% and complete QA at checkout
Specifications
Display: LCD, 4 1/2 digit, 0.48” high Range: Charge: .0001x10-10 to 12.000x10-7C Current: .0001x10-11 to 12.000x10-1A Overrange: Indicated by 1 followed by blanks Accuracy: ±0.1% of reading + 1 digit Linearity: ±0.1% + 1 digit or precision of reading Leakage: <5x10-15A Calibration: coulomb scale to ±0.1% Input Impedance: >1x1014 ohm Temp. Coefficient: 0.2%/EC Bias: internal electronic bias supply; -150V, -300V, OFF, +300V, +150V, selectable on front panel, bias voltage may be read on display External Bias: Binding posts for banana plugs Input: Triaxial BNC-F with cap & chain Power: rechargeable 12V 1.2 AH gel-cell battery, voltage may be read on display, low battery indicator, 120VAC UL listed wall- mount charger provided (240VAC optional) Current Drain: 57mA Dimensions: 7” Wide x 10.75” High x 11.5” Deep Weight: 14.5 lbs
|