TOPEX SRT 100 Superficial Radiotherapy System is designed to meet today's needs of the medical practitioner for the treatment of skin cancer.
TOPEX has developed a totally unique SRT product design through consultation with the medical community and expertise from within the x-ray therapy industry. This new design provides a system with compact size, mobility, automated set-up controls, modular component electrical assembly platforms and a wide range of motion for patient treatment flexibility. The SRT 100 represents a culmination of "bottom-up" product design effort resulting in significant improvements compared to older bulky component system designs.
Advanced System Design for Treating Skin Cancer:
• Large Range of Motion The unique positioning arm design has both horizontal and vertical range of motion.
• Sitting or Recumbent Patient Positioning Flexible movement allows for clinical treatment with patients in sitting or recumbent positions.
• Compact Mobile Design The unique mobile design requires only a 29” x 30" footprint saving space while providing treatment location flexibility.
• Pre-Treatment "RAD" Check This unique feature verifies dose delivery required before the patient is treated, and is displayed on the control console.
• Automatic Warm-up There are four-programmed x-ray tube warm-up sequences that are selected based on the time of last exposure.
• Automatic Al Filter Change (Patent Pending) Upon kV mA selection, the system automatically changes the Al Filter eliminating the need for filter storage and manual procedures.
• Precise Dose Delivery There are two x-ray beam energies used specifically for the treatment of skin cancer. The x-ray tube focal spot is 1mm x 1mm and reduces the penumbral effect providing a more uniform treatment field.
• Replaceable Visual Applicator Safety Contact Shields The system includes three applicators: 1.5cm, 3cm and 5cm. Each applicator safety contact shield is replaceable and provides for 1.27cm visual verification and clinical margin clearance of the treated area.
Specifications
Generator: Type- Constant Potential HV, Input Line- 120 - 230 VAC
X-Ray Tube: Metal Ceramic, Water-cooled, Tungsten Target, End grounded Rating= 120kV/10mA, 40kV/30mA, 1200 watts continuous dissipation Focal Spot Size= 1mm x 1mm
Base Unit Assembly: Basic Space Requirements: 29” x 30” Vertical Travel: up to 68” Horizontal Arm Travel: 17” to 61” from center X-Ray Tube Movement: V&H 180 degrees Integrated Modular Components: Input power, HV generator, Heat exchanger
Operator Control Console: Can be located up to 100’ (30 meters) from base unit Service mode for system set-up and calibration - key entry
Two Treatment Techniques: 100kV @ 8mA, 2.1mm Al. HVL or 70kV @ 10mA, 1.3mm Al. HVL X-Ray output is 460 cGy @ 15cm SSD
Automatic Filter Change (Patent Pending): 2.1mm Al. HVL / 1.3 mm Al. HVL / Pb X-Ray Block
Automatic Warm-up Procedures: Automatically activated from the time of last exposure Four pre-programmed sequences Lead blocker automatically placed over x-ray tube support
RAD Check: Direct radiation measurement of output and pre-treatment verification
System Weight: 350 lbs
Accessories:
The SRT 100 includes all the various sized Applicators ranging from 1.5cm to 10.0cm, an Applicator Replacement TIP Kit, a Service Spare Fuse/Hardware Kit, a Cable Retractor Installation Kit and a set of Physicist Calibration Fixtures.
Replaceable Safety Contact Shields: Applicator size specific. Visibility of treated area.
Maximum Field Size: 10cm
Maximum Energy: 100kV
|