Designed for 3M Type Cesium-137 Sources
The Wang-Front Loading Applicator is an intracavitary afterloading device utilizing three Cesium-137 sources in a “T” configuration. The three 3M Cesium-137 (Cs-137) tube sources each have a diameter of 3.1 mm and length of 20.0 mm. The sources are arranged in a “T” configuration, with two tandem sources oriented along the longitudinal axis of the cylinder and one apex source perpendicular to them.
This source geometry provides a homogenous dose distribution over the apex and lateral walls of the applicator. Computer point dose calculations, generated by a traditional brachytherapy program and Monte Carlo program, agreed within standard errors of measurements obtained by TLD in a plastic phantom. The vaginal cylinder is 30.0 mm in diameter and 70.0 mm in length. The apex of the applicator is semi-spherical. The cylinder is sealed to prevent fluid leakage into the source cavities.
Set Consists of: (1) Wang Applicator (1) Bucket-Type Source Carrier (1) Flexible Source Carrier (2) Tandem End Caps (1) 15cm Stainless Steel Ruler (1) 12” Hemostat (1) Stainless Steel Sterilization Tray (1) Wooden Carrying Case
Instruction Manual Sterilization: Autoclave or Gas
|