View PDF Catalog   

Closeout Items   

RAD-60 Personal Electronic Dosimeter, Rem Display

RAD-60 Personal Electronic Dosimeter, Rem Display
RAD-60 Personal Electronic Dosimeter, Rem Display
Item Number :
343-600

RAD-60 Personal Electronic Dosimeter, Rem Display.

Details

Open Section I: Dosimeters, Monitors, Meters, Check Source & Lead Cases(PDF) to view related products. (Opens in a new tab)

RAD-60 Personal Electronic Dosimeter, Rem Display.

The RAD-60 Personal Alarm Dosimeter is a precise radiation measuring instrument for reliable detection and registration of radiation in order to ensure the personal safety of the user. It is suitable for a broad range of everyday radiation monitoring purposes in stand alone conditions.

• Enhanced EMI immunity
• Improved wear-out and decontamination properties
• Individual Personal Alarming Dosimeter
• Digital display for integrated dose or alternatively dose rate
• User selectable alarm levels for both dose and dose rate
• Detector system utilizes high quality energy compensated Si diode and advanced mathematical dose rate linearization
• Splash-proof mechanical construction, high impact plastic with strong pocket clip

 

Applications
• Civil Defence, Rescue Operations
• Customs Operations, Military Forces
• Industrial radiography, Nuclear Medicine


Specifications
Radiation Detected: Gamma and X-rays
Detector Type: Energy-Compensated Si-Diode Measurement Range
     Dose: 1 µSv - 9.99 Sv or 0.1 mrem-999 rem
     Dose Rate: 5 µSv/h - 3 Sv/h or 0.5 mrem/h - 300 rem/h
Calibration: Better than ±5 % (Cs-137, 662 keV at 2 mSv/h), Hp(10)
Energy Response: Hp(10),  55 keV - 3 MeV, better than ±25%, up to 6 MeV, better than ±35%
Dose Rate Linearity: Better than ±15%, up to 3 Sv/h (300 rem/h)
Alarm Thresholds: Seven preset values each for integrated dose and dose rate, manually selectable by push button.
Font Panel Push-Button Functions:
     Toggle between dose and dose rate display Switch ON/OFF
     Chirp ON/OFF
     Reset Integrated Dose
     Change Alarm Thresholds
     Activate Battery Test
Audible Alarms: Eight separate alarms, sound level typically better than 85 dBa at 30 cm
     Integrated dose
      Dose rate
      Dose overflow
      Dose rate overflow at 3 Sv/h or 300 rem/h Low battery 1 and 2
      Defect
Power Supply: One AAA alkaline cell, typical life is 1800 hrs in background field (dose mode)
Reader Communication: By infrared through bottom part; by using ADR-1 Reader Head in combination with PC Software
Temperature Range: -4° - 122° F (-20 to +50°C) operational, humidity up to 90% RH, non-condensed
Dimensions: 3.07" x 2.63" x 0.86" (78 x 67 x 22 mm)
Weight: 2.82 oz (80 g) including battery