Exradin A14SL Slimline Microchamber, 0.015cc
300-250-A14SL
Exradin A14SL Slimline Microchamber, 0.015cc.
Exradin A14SL Slimline Microchamber, 0.015cc.
Exradin A14SL Slimline Microchamber, 0.015cc.
The Exradin micro ion chambers have a superior small-field dosimetry to assess pinpoint radiation in IMRT, stereotactic, orthovoltage, x-rays and superficial skin therapy.
The Model A14SL is capable of measuring extremely small field sizes of 4 mm by 6 mm, allowing for exceptional spatial resolution and exact pinpoint beam profile characterization. This helps assess radiation fields during IMRT and stereotactic radiosurgery.
Specifications
Collecting Volume: 0.015cc
Centroid of Collecting Volume - from exterior tip of shell: 2.39 mm
Outside of Shell Collecting Diameter: 6.35 mm
Inside Diameter of Shell Collecting Volume: 4.0 mm
Shell Wall Thickness: 1.1 mm
Collector Diameter: 0.3 mm
Collector Length: 1.5 mm
Shell/Entry Window, Collector and Guard Material*: Shell Only A
Nominal Air Kerma Calibration Factor†: 1.9E + 9 Gy/C
Recommended Polarizing Voltage: 300 V
Nominal Leakage: ±10 x 10-15 amp
Maximum Polarizing Voltage: 1000 V
Waterproof: Yes
Included Build-Up Cap: None
* Material: A – C552 Shonka air-equivalent plastic P – D400 polystyrene-equivalent plastic T – A150 Shonka tissue-equivalent plastic.
† Nominal calibration factor for Co-60 at 22° C